|
|||||||||||||||
|
Minimum ten-year follow-up of the HA-coated Esop
The implants The stability of the insert is ensured
by the circumferential clamping action of the cup around the
peripheral cylindrical portion of the insert. This design
counteracts any tendency of the insert to tilt
8
. The thickness of the cup shell is
limited to 2.5 mm, so as to allow a maximal thickness of PE
to be inserted. The femoral heads used were 22.2-mm
diameter in 145 cases, and 26-mm diameter in six cases. Like
the thin shell, these small heads allowed a greater
thickness of PE to be used. The PE insert was always at
least 10 mm thick. On the femoral side, a PSM stem
(HA-coated in the metaphyseal portion) was used in 147
cases; the remaining four cases were managed with a variety
of cementless implants. Materials and method : Twelve patients had died, and 16 were
lost to follow-up. Fourteen patients were reoperated on: one
for infection, three for dislocation, four for failure of
the stem to obtain primary stability, four for metallosis
caused by the deterioration of the titanium femoral head,
and two for cup loosening. This left 109 patients for review
in the present study, at a follow-up of eleven years
post-arthroplasty in 61 cases, and of ten years in 48 cases.
For the clinical assessment, the MDA scoring system was
used. The radiographic assessment was concerned with the
detection of lucent lines, cysts, and evidence of loosening,
on the a.p. and the lateral films. Wear was measured on the
A/P hip films using the Charnley method, 5
taking magnification into account. Cumulative survival out
to eleven years could be calculated, with the Kaplan-Meier
method, for the 151 hips that had undergone replacement. Results In the 109 cases that were followed
up, the MDA pain score was 6 in 95 cases, 5 in ten cases,
and 4 in five cases. The MDA mobility score was 6 in 98
cases, 5 in seven cases (who had slight stiffness), and 4 in
four cases (whose hips were stiff). The MDA function score
was 6 in 93 cases (patients walking normally), 5 in seven
cases (slight limp), and 4 in nine cases (limp). Seven of
the last-mentioned nine patients were DDH cases. Radiography showed a total absence of
any periacetabular cysts. In the femur, there were 15 cysts,
all of which were confined to the greater trochanter (Figure
3); there were no extensive cysts and no extensive lucent
lines. At the level of the cup,
non-progressive lucent lines of less than 3 mm width were
seen in Zone III in seven cases (6.2%). There were no
extensive lucent lines. Wear was 0 mm in eleven cases, 0.5 mm
in 31 cases, 1 mm in 42 cases, 1.5 mm in 16 cases, 2 mm in
seven cases, and 2.5 mm in one case. Mean wear was 0.90 mm.
The mean annual wear rate in this series was 0.82 mm/year.
Mean cumulative survival was
calculated, with the Kaplan-Meier method, for the 151 hips
in the series. It was calculated out to eleven years, since
78 of the hips were still "alive" by the end of that time.
The seven stem failures were not included in the analysis,
since the study was concerned with cup survival. Also,
following the policy of the Scandinavian registries, it was
decided to apply the term 'revision' only to reoperations
involving the exchange of at least one of the two major
components in direct contact with the host bone (i.e. the
cup and/or the stem), while the exchange of an acetabular
insert or a femoral head would not be considered to be a
'revision'. There were three failures involving
the Esop cup, one each in the first, the fifth, and the
ninth year post-operatively. Two were revisions for
'intercurrent' reasons (one deep infection in the first
year, the other as a result of an accidental fracture at
nine years). Only in the case of the cup that needed
revising at five years was failure of the implant fixation
to blame. Thus, when analyzing cup failure as such, the rate
was 1.98% (3/151 cups). This figure comprises the two
failures not attributable to the implant as such, and one
failure of implant fixation. The failure rate of the HA
fixation of the Esop cup at ten years was thus 0.66% (1/151
cups). When these data were used to plot the survival curves, the cumulative survival rates at eleven years, using the different endpoints, were seen to be - cup replacement (all causes): 96.95 % ± 0.0269 Esop cup mechanical failure: 99.18% ± 0.0159
|
|||||||||||||||
|
|||||||||||||||
|
Discussion : The absence of macrophage-filled cysts
around the cup provided evidence of the firm seating of the
PE insert in the cup shell. The insert fixation system by
circumferential clamping in the cup proved efficient in the
clinical study, as it had done previously in a retrieval
study. 10
In that study, no macroscopic or microscopic evidence of
backside deterioration of the inserts had been found after
more than ten years post-operatively. These results were
certainly due, in some measure, to the absence of
microfretting between the insert and the shell thanks to the
stable seating of the insert, and to the great thickness of
the PE. Another factor was the size of the femoral heads
used. Most of the heads were 22-mm diameter ones, and this
size is known to produce little wear debris and,
consequently, little macrophage reaction. The findings in
this study thus bear out Charnley's low-friction theory.
Lucent lines around the cup were rare
(6.2%) and confined to Zone III. It is interesting to recall
that, in a comparative study in 1995, 6
we found the non-HA-coated Esop cups to be associated with a
40% rate of Zone III lucent lines. Many authors 2
11
121317
have described the barrier to the migration of PE wear
particles provided by the intimate contact, without any
interposed fibrous tissue, between the host bone and a
circumferential HA coating on the implant. It would appear
that, in the case of the implants used in the present study,
this barrier effect resulted in any femoral cysts being
confined to the greater trochanter. Linear PE wear was slight (0.8
mm/year). While Bloebaum et al 3
have voiced concern over possible third-body wear by
dislodged HA particles becoming embedded in the PE bearing
surface, this study showed that a good quality HA coating
does not constitute such a hazard. While Nashed16
, Bankson1
and Hernandez14
with metal backed acetabular prostheses were refuted by the
publications of Markel15
and of Callaghan4
found low wear rates. Wear is a multifactorial phenomenon,
with PE quality, PE thickness, femoral head quality,
patient's level of physical activity, component positioning,
and other factors each playing a role. Under the conditions
of use described in this study, there appears to be no cause
for concern about accelerated PE wear.
|
|||||||||||||||
![]() More information |
|||||||||||||||
|
Conclusion :
|
|||||||||||||||
|
Surgical procedure - Clinical data - Quality requirement - The inventor - Contact - International distribution |